Description
As a necessary choice for quantitative analysis of molecular biology, Real-time PCR system has been widely used in various fields such as scientific research, clinical detection and diagnosis, quality and safety testing, and forensic applications.
- Up to 6 fluorescence detection channels allowing multiplex PCR.
- Effectively reduce multi-color crosstalk and edge effect, no ROX correction required.
- New optical scanning detection system
- Innovative scanning method and time-resolved signal separation technology
- Unique edge temperature compensation technology
Specifications
| Model | CA-PCR-Q96-5 |
CA-PCR-Q96-6 |
| Sample capacity | 96 | |
| Reaction volume | 10-50 μl | |
| Thermal cycle technology | Peltier | |
| Temperature control system | ||
| Max. Heating/Cooling rate | 6.0° C/s | |
| Heating temperature range | 4 – 100 °C | |
| Temperature accuracy | ± 0.2°C | |
| Temperature uniformity | ±0.2℃ @60℃ , ±0.3℃ @95℃ | |
| Temperature gradient setting range | 30–100°C | |
| Temperature gradient difference setting range | 1 – 36°C | |
| Detection system | ||
| Excitation light source | 4 monochrome high efficiency LEDs | 6 monochrome high efficiency LEDs |
| Detection device | PMT | |
| Detection mode | Time-resolved signal separating technology | |
| Excitation/detection wavelength range | 455-650nm/510-715nm | |
| Fluorescent channels | 4 channels | 6 channels |
| Supported dye | FAM/SYBR Green, VIC/JOE/HEX/TET, ABY/NED/TAMRA/Cy3, JUN, ROX/Texas Red, Mustang Purple, Cy5/LIZ | |
| Sensitivity | Single copy gene | |
| Resolution | 1.33 folds copy number difference can be distinguished in single-plex qPCR | |
| Dynamic range | 10 orders of magnitude copies | |
| Package Dimension (W*D*H)(mm) | 960*660*750 | |
| G.W.(kg) | 47 | |