Patient Monitor – 13 Type Arrhythmic Analysis Including ECG and Respirational, CA-PM-15D

  • Available for neonatal monitoring
  • 15″ color TFT LCD display
  • Anti-ESU, anti- defibrillator
  • Standerd:ECG, Respiration, NIBP, SpO2, Pulse Rate, Temp.
  • Networking capacity
SKU: 14012 Category:

Description

  1. 15″ high resolution color TFT LCD display
  2. 13 Type Arrhythmic Analysis, Multi_Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable
  3. Efficient resistance to interference of defibrillator and electrosurgical cautery
  4. SPO2 can testing for 0.1% Weak
  5. RA-LL impedance Respiration
  6. Trend Coexist Display
  7. OxyCRG DyNamic View Display
  8. Bed to Bed view Display
  9. Networking capacity
  10. Options of printing, IBP, VGA and EtCO2
  11. UP to 4 hours working capacity of built_in rechargeable battery
  12. Large volume of tabular and graphic trends information storage and easy to recall
  13. Anti-ESU, anti- defibrillator
  14. Capture dynamic waveforms.

Specifications

ECG Lead Mode: 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V)
Lead selection: I, II, III, avR, avL, avF, V,
Waveform: 2 ch
Lead mode: 3 Leads ( R, L, F or RA, LA, LL)
Lead selection: I, II, III,
Waveform: 1 ch
Gain: ′2.5mm/mV, ′5.0mm/mV, ′10mm/mV, ′20mm/mV,  auto
HR and Alarm
Range Adult: 15 ~ 300 bpm
Neo/Ped: 15 ~ 350 bpm
Accuracy : ± 1% or ± 1bpm,which great
Resolution: 1 bpm
Sensitivity: > 200 (uV P-P)
Differential Input Impedance: > 5 MΩ
CMRR Monitor: > 105 dB
Operation: > 105 dB
Diagnosis: > 85 dB
Electrode offset potential: ±300mV
Leakage Current: < 10 uA
Baseline Recovery: < 3 S  After Defi.
ECG Signal Range: ±8 m V ( Vp-p )
Bandwidth
Surgery: 1 ~ 15 Hz
Monitor: 0.5 ~ 35 Hz
Diagnostic: 0.05 ~ 100 Hz
Calibration Signal: 1 (mV p-p),  Accuracy:±5%
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC,
TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm: Available
Review: Available
RESP Method: Impedance between R-F(RA-LL)
Differential Input Impedance: >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
Base line Impedance Range: 0 – 2.5  KΩ
Bandwidth: 0.3 ~ 2.5 Hz
Resp. Rate Measuring and Alarm Range
Adult: 0 ~ 120 rpm
Neo/Ped: 0 ~ 150 rpm
Resolution: 1 rpm
Accuracy : ±2 rpm
Apean Alarm: 10 ~ 40 S
NIBP Method: Oscillometric
Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)