Analysis system |
Working mode |
Discrete/random access |
Test speed |
test speed 180T/H (without ISE) |
Test principle |
Absorbance photometry, Turbidimetry |
Methodology |
End-point, Fixed-time, Kinetic, Single/Dual reagent chemistries, monochromatic/bichromatic Linear/non-linear multi-point calibration |
The longest reaction time |
15 minutes |
Minimum reaction volume |
150μl |
Cuvettes material |
plastics (quartz glass can be selected), 81 cuvettes have the function of automatically deducting reagent and sample blanks |
Linear range of absorbance |
0-3.8Abs |
Halogen light source using time |
≥ 2000 hours |
Wavelength: |
340nm、405nm、450nm、510nm、546nm、578nm、630nm、670nm |
Refrigeration system |
water medium uniform refrigeration technology |
Reagent storehouse temperature |
4-16℃ |
Reaction temperature |
37±0.2℃ |
Temperature fluctuation |
±0.1℃ |
Sampling and mixing system |
Sample capacity |
2-30μl, 0.1μl increasing |
Sample probe |
Sharing with the reagent probe,the inner and outer walls are highly polished,with the function of liquid level detection, tracking and collision protection,sample probe matching special cleaning liquid,prevent cross contamination |
Reagent capacity |
20-300μl, 0.1μl increasing |
Reagent probe |
Sharing with the sample probe |
Reagent bottle |
Reagent bottles with volume of 15ml、 20ml、 30ml、 50ml、 60ml 、 70ml can be used, the dead volume is less than 1ml |
Mixing bar |
Independent mixing bar, special material, surface treatment using Teflon, not hanging liquid, thoroughly clean before and after mixing to prevent cross contamination |
Reagent/sample Handing |
Reagent/sample tray |
60 positions for reagents and 30 positions for samples in refrigerated compartment (4~16℃) |
Reagent volume |
20-500μl,step by 1μl |
Sample volume |
2~100μl,step by 0.1μl |
Reagent/Sample probe |
Liquid level detection,collision protection and inventory checking |
Probe cleaning |
Automatic washing for both interior and exterior Carry-over < 0.1% |
Calibration and quality control |
Quality control type |
real time quality control, daily quality control, day to day quality control |
Quality control charts |
L-J, Cumulative, Twin Plot |
The quality control rules |
arbitrarily formulated by default to Westguard multiple rules |
The calibration type |
linear and nonlinear. Logit-4P, Logit-5P, Spline, exponent, polynomial, factor method |
Calibration |
Automatically check the calibration curve and select the best calibration type of the fitting degree automatically |
Operating System |
English operating system: |
Windows2000、windows XP、windows7、windows8、windows10 |
Report printing |
Make print format and content freely, provide several common formats |
user management |
Special user management password, rational distribution of user operation permissions |
Interface |
RS-232 standard interface, support for intranet, remote print report |
Databases |
Optional various databases, autosave, automatic backup, save all kinds of data permanently |
Real-time monitoring of sample tray、reagents tray、reacting tray;Real-time display of reaction temperature, reagent allowance, reaction curve, calibration curve and quality control chart |
Check and judge reaction endpoint, reaction linear interval, substrate exhaustion, reagent blank absorbance, etc. |
Language |
English, Chinese |
Reviews
There are no reviews yet.