Description
Features
- Dual refrigeration system. (Excluding ICB-MS175-H,ICB-MS175T-H,ICB-MS175-HL,ICB-MS175T-HL,ICB-MS175-L)
- Imported compressor with self-developed compressor cooling system, which extends the service life of the compressor.
- Imported humidity sensor, ensuring high accuracy of humidity.
- PID control for temperature and humidity is more accurate and has an easy menu operation.
- Over-temperature alarm system: automatically stops working when the limit temperature is reached and sends out an audible and visual alarm, ensuring experiments run safely.
- Large LCD screen/touch screen(T-H series) to display more data at the same time.
- Newest multi-segment programmable control technology for temperature and humidity, with high precision. With multiple programs and multiple cycles, each cycle is divided into 30 segments, each segment contains 99 hours and 99 minutes of cycle steps, which can happily meet almost all the complicated experiment processes.
- Equipped with a JAKEL circulation fan, a uniquely designed air duct for good air circulation, ensuring good temperature uniformity inside.
- Equipped with a built-in stainless steel water tank at the bottom of the equipment to supply water by pump for the humidification tank.
- Equipped with a spare temperature control, which ensures the product works normally even if the main temperature.control failed(for heating).
- Equipped with a printer that can record and print the operating parameters at any time.
- Mirror stainless steel chamber, adjustable shelf.
- The chamber is equipped with a power socket and UV lamp for sterilization.
- Double door design. Tempered glass inner door for clear observation. The outer door adopts a magnetic seal design with good sealing performance.
Safety Device
- Over-temperature protection
- Compressor overcurrent protection
- Over-compressing protection
- Water shortage protection
- Heater overheat protection
- Audible and visual alarm system
Specifications
| Model | ICB-MS175-H/ ICB-MS175T-H ICB-MS275-H / ICB-MS275T-H ICB-MS375-H / ICB-MS375T-H ICB-MS475-H / ICB-MS475T-H ICB-MS800-H / ICB-MS800T-H ICB-MS1075-H / ICB-MS1075T-H |
ICB-MS175-HL/ ICB-MS175T-HL ICB-MS275-HL / ICB-MS275T-HL ICB-MS375-HL / ICB-MS375T-HL ICB-MS475-HL / ICB-MS475T-HL ICB-MS800-HL / ICB-MS800T-HL ICB-MS1075-HL / ICB-MS1075T-HL |
ICB-MS175-L ICB-MS275-L ICB-MS375-L ICB-MS475-L ICB-MS800-L ICB-MS1075-L |
| Temperature & Humidity | Temperature & Humidity & Light | Temperature & Light | |
| Temperature Range (℃) | 0~65 | Without Illumination:4~60 With Illumination:10~60 |
|
| Temperature Fluctuation | High±0.5℃、Low±2℃ | ||
| Temperature Uniformity (℃) | ±0.5 | ||
| Humidity Range (RH) | 30~95% | None | |
| Humidity Stability (RH) | ±3%RH | ||
| Temperature Resolution (℃) | 0.1 | ||
| Illumination (LX) | None | 0~6000 adjustable | |
| Illumination Difference (LX) | ≤±500 | ||
| Timing Range | 1~99 hours/period | ||
| Temp.and Humidity adjusting | Balance temperature and humidity adjusting | Balance temperature adjusting | |
| Cooling System | Imported compressor | ||
| Sensor | PT100 Capacitance sensor |
PT100 | |
| Ambient Temperature | RT+5~30℃ | ||
| Power Supply | AC 220V±10% ,50HZ | ||
| Power Rating (W) | 1400/1950/2600/2800/3000/3200 | 1650/2200/2700/2900/3100/3300 | 1300/1750/2400/2600/2700/2800 |
| Chamber Volume (L) | 175L, 275L, 375L, 475L, 800L, 1075L | ||
| Chamber Size (W×D×H)mm | 450×420×930 580×510×935 590×550×1160 700×550×1250 965×610×1370 950×700×1600 |
||
| Shelf | 3 | ||
| Printer | Yes | ||
| Safety Device | Compressor overheating and over-pressure protection, Fan overheating protection, Over temperature protection, Overload protection, water shortage protection |
||