Description
- 21 parameters + 3 histograms
- 10.4 inches color touch screen display
- Up to 60 tests per hour;
- 3-part differentiation of WBC
- Double chambers for WBC and RBC/PLT
- External thermal printer
- Barcode scanner for option
- Storage ≧200,000 sample results
SPECIFICATIONS
21 Parameters | WBC,LY#,MID#,GR#,LY%,MID%,GR%,RBC,HGB,HCT,MCV,MCH,MCHC,RDW-SD,RDW-CV,PLT,MPV,PDW,PCT,P-LCR,P-LCC |
Histogram | Histogram for WBC,RBC,PLT |
Method | Coulter principle |
Colorimetry hamoglobin determination | |
Throughput | 60 tests/hour |
Display | 10.4″ color touch screen display |
Language | Chinese, English, Portuguese, Hindi, Spanish, French, Thai, Vietnamese |
Sampler Volume | Predilute 20μl |
Whole blood 10μl | |
Intelligent reagent management function | Reagent information and residual display |
Carryover | WBC≤1.5% |
RBC,HGB<1.0% , PLT≤2% | |
Input and output | 1 RS232 , keyboard , Mouse |
4USB,LAN | |
Printout | External thermal printer |
Option | Barcode scanner, LIS |
Operating Environment | Temperature 10-35℃ |
Humidity≤90% | |
Data storage | Over 200,000 sample results (with histograms) |
Power requirement | AC 100-240V |
50/60HZ , 300VA | |
Demension | Instrument: 354mm(H)*430mm(W)*440mm(D) |
Package: 520mm(H)*570mm(W)* 600mm(D) | |
Weight | Net :17kg |
Gross:21. 8kg |
Performance
Parameter | Linearity range | Precision(CV%) |
WBC(10 9/L) | 0.0-100.0 | 2.0(4.0-10.0) |
RBC(1012/L) | 0.00-9.99 | 1.5(3.50-5.50) |
HGB(g/L) | 0-300 | 1.0(110-160) |
PLT(109/L) | 0-1000 | 5.0( 100-300) |
ACCESSORIES
Reviews