Fully auto Biochemistry Analyzer, CA-BA-A-200

  • 180 tests per hour
  • 30 positions for samples and 60 positions for reagents
  • Sample capacity: 2-30μl, 0.1μl increasing
SKU: CA-BA-A-200 Category:

Description

  1. Discrete, random access, fully automated
  2. Up to constant 180 tests per hour
  3. Optional for external bar code reader
  4. 30 positions for samples and 60 positions for reagents
  5. Automatic probe cleaning, liquid level detection, collision protection
  6. Reversed optic system with 8 wavelengths
  7. Refrigerated reagent and sample compartment

Specifications

Analysis system
Working mode Discrete/random access
Test speed test speed 180T/H (without ISE)
Test principle Absorbance photometry, Turbidimetry
Methodology End-point, Fixed-time, Kinetic, Single/Dual reagent chemistries, monochromatic/bichromatic Linear/non-linear multi-point calibration
The longest reaction time 15 minutes
Minimum reaction volume 150μl
Cuvettes material plastics (quartz glass can be selected), 81 cuvettes have the function of automatically deducting reagent and sample blanks
Linear range of absorbance 0-3.8Abs
Halogen light source using time ≥ 2000 hours
Wavelength: 340nm、405nm、450nm、510nm、546nm、578nm、630nm、670nm
Refrigeration system water medium uniform refrigeration technology
Reagent storehouse temperature 4-16℃
Reaction temperature 37±0.2℃
Temperature fluctuation ±0.1℃
Sampling and mixing system
Sample capacity 2-30μl, 0.1μl increasing
Sample probe Sharing with the reagent probe,the inner and outer walls are highly polished,with the function of liquid level detection, tracking and collision protection,sample probe matching special cleaning liquid,prevent cross contamination
Reagent capacity 20-300μl, 0.1μl increasing
Reagent probe Sharing with the sample probe
Reagent bottle Reagent bottles with volume of 15ml、 20ml、 30ml、 50ml、 60ml 、 70ml can be used, the dead volume is less than 1ml
Mixing bar Independent mixing bar, special material, surface treatment using Teflon, not hanging liquid, thoroughly clean before and after mixing to prevent cross contamination
Reagent/sample Handing
Reagent/sample tray 60 positions for reagents and 30 positions for samples in refrigerated compartment (4~16℃)
Reagent volume 20-500μl,step by 1μl
Sample volume 2~100μl,step by 0.1μl
Reagent/Sample probe Liquid level detection,collision protection and inventory checking
Probe cleaning Automatic washing for both interior and exterior Carry-over < 0.1%
Calibration and quality control
Quality control type real time quality control, daily quality control, day to day quality control
Quality control charts L-J, Cumulative, Twin Plot
The quality control rules arbitrarily formulated by default to Westguard multiple rules
The calibration type linear and nonlinear. Logit-4P, Logit-5P, Spline, exponent, polynomial, factor method
Calibration Automatically check the calibration curve and select the best calibration type of the fitting degree automatically
Operating System
English operating system: Windows2000、windows XP、windows7、windows8、windows10
Report printing Make print format and content freely, provide several common formats
user management Special user management password, rational distribution of user operation permissions
Interface RS-232 standard interface, support for intranet, remote print report
Databases Optional various databases, autosave, automatic backup, save all kinds of data permanently
Real-time monitoring of sample tray、reagents tray、reacting tray;Real-time display of reaction temperature, reagent allowance, reaction curve, calibration curve and quality control chart
Check and judge reaction endpoint, reaction linear interval, substrate exhaustion, reagent blank absorbance, etc.
Language English, Chinese