Graphite Furnace System, CA-GRD-3202

  • Heating steps: 9 steps
  • Temperature range (nominal Temperature): 20℃~3000℃
  • Slope heating time: 0~999s
  • Heating holding time: 1s~999s (the sum of both time should be less than 999 seconds)
  • Inert gas needed: argon, pressure of entry larger than 0.3Mpa
  • Cooling water: tap water or cycling water, flow rate no less than 2L/min
SKU: CA-GRD-3202 Category:

Description

GRD-3202 graphite furnace system is a system specially used for the quantitative analysis of trace elements in substances by graphite furnace atomic absorption method. It is an important optional accessory for SP-IAA320 atomic absorption spectrophotometer.

Specifications

Model GRD-3202
Heating steps 9 steps
Temperature range (nominal Temperature) 20℃~3000℃
Slope heating time 0~999s
Heating holding time 1s~999s (the sum of both time should be less than 999 seconds)
Inert gas needed Argon, pressure of entry larger than 0.3Mpa
Cooling water Tap water or cycling water, flow rate no less than 2L/min
Output LCD
Power source 220V±22V, 50Hz±1Hz, 3A 220V±22V, 50Hz±1Hz, 30A
Power 5.1kW for 220V on about 2700℃
Ambient temperature +10℃~+30℃
Relative humidity Less than 85%
Instrument rating power 7.2kW
Gas pressure alarm, furnace overheated alarm
Interface with Atomic absorption spectrophotometer, autosampler and RS232
High-power temperature heating function (1000℃~2700℃)
Test sample together with atom absorption equipment. Cd ≤1×10^-12g; Cu ≤1×10^-10g
Package Dimension (W*D*H)(mm) 730*660*910

Reviews

There are no reviews yet.

Be the first to review “Graphite Furnace System, CA-GRD-3202”

Your email address will not be published. Required fields are marked *