Disintegration apparatus, CA-DITT-2

  • Temperature prefabrication range :R.T.:~50℃
  • Display resolution: 0.1℃.
  • Number of lifting baskets:  2 sets
SKU: CA-DITT-2 Category:

Description

 

Disintegration time limit tester is a drug test instrument to test the disintegration time limit of tablets, capsules and pills in accordance with Chinese Pharmacopoeia.

  • Double-basket working position system measures disintegration time limit.
  • 2 sets of baskets, which can conduct independent disintegration experiment respectively.
  • Electronic temperature sensor can display and monitor various points in water bath case and the temperature in the beaker. High-precision digital electronic sensor is adopted to ensure the high precision and high accuracy of water bath temperature without the need for calibration.
  • The instrument controls the bath temperature at 37.0℃ automatically (prescribed by Pharmacopeia). It can also reset the preset temperature again at any time.
  • The instrument sets the lifting time of baskets to be 15min automatically, which can also be reset at any time.
  • It adopts the computer control technology which is centered on SCM to control two performance parameters including bath temperature and working time intelligently.
  • The instrument can monitor bath temperature, give over-temperature alarm and protect itself automatically.
2 Sets of Baskets 2 Sets of Baskets
Electronic Temperature Sensor Electronic Temperature Sensor

SPECIFICATIONS

Model CA-DITT-2
Temperature prefabrication range room temperature: ~50℃
display resolution: 0.1℃.
Temperature control accuracy ±0.3℃
Timing prefabrication range 1min~900min
Time control accuracy ±0.5min
Work noise <60db
Number of lifting baskets 2 sets
Lifting frequency of basket (30~32) times/min.
Lifting distance of basket (55±1) mm
Minimum spacing from screen to beaker bottom 25mm±2mm
Mesh size standard configuration 2mm, (optional for 0.425mm and 1mm bore diameters)
Power supply 220V50Hz
Total power 600W
Overall dimensions L*W*H 400mm*320mm*440mm