Gas Chromatograph, CA-GC-I126N

  • Detector: FID, TCD, ECD and FPD are optional
  • Standard dual-channel data acquisition card
  • Self-check function & Fault auto-identification function
  • Auto control Split flow/split ratio
SKU: CA-GC-I126N Category:

Description

1.The host machine adopts a 7-inch color touch screen with friendly man-machine interface.
2. Computer back control and host touch screen to achieve synchronous bidirectional control.
3. The multi-core, 32-bit embedded hardware system ensures the reliable operation of the instrument.
4. Carrier/hydrogen/air flow (pressure) digital display.
5. Gas shortage alarm protection function; Heating control protection function (when opening the door of the column box, the motor of the column box fan and the heating system will shut down automatically).
6. Split flow/split ratio can be automatically controlled to save carrier gas.
7. Configure automatic sampler installation and positioning interface to match automatic sampler of various specifications.
8. Data acquisition is a standard dual-channel data acquisition card with a sampling time of 50ms.
9. Using logarithmic amplification plate, detection signal no cut-off value, synchronous external trigger function, can be started by external signals (automatic sampler, thermal analyzer, etc.) at the same time the host and workstation.
10. It has perfect system self-check function and fault automatic identification function.
11. With eight external event extension function interface, can be selected with a variety of control valves, and according to their own set time sequence work.
12. The external link mode is network port connection (RJ45), which provides convenience for remote control of data.

Specifications

Column temperature box

  • Temperature range: 5℃ ~ 400℃ at room temperature
  • Temperature control accuracy: ±0.1℃
  • Program heating: Stage 9/10 platform
  • Total program time: 9999.9 min
  • Maximum heating rate: 60℃/min

Sampler

  • Temperature range: 7℃~ 420℃ at room temperature
  • Fill column, split/split sampler can be installed
  • Working under constant pressure mode
  • No more than three independent sampling systems are connected simultaneously

Detector

  • A maximum of 2 units are installed simultaneously. FID, TCD, ECD and FPD are optional.
  • Flow/pressure is displayed directly on the screen.
  • Temperature setting: Max. 420℃
  • Hydrogen flame ionization detector (FID)

High voltage switch control

Baseline signal display

Ignition coil control

Detection limit: ≤ 3×10-12 g/s

Sample: N-hexadecane (minimum detection quantity: 3pg/s)

Dynamic range: 107

  • Thermal conductivity detector (TCD) 

Bridge voltage switch control

Bridge current setting: 0 ~ 220 mA

Sensitivity: 5000 mV×mL/mg

Dynamic range: 105

  • Electronic capture detector (ECD) 

Radiation source: Ni63

Detection limit: ≤8×10-14 g/s

Sample: R 666 (minimum detection quantity: 80FG/s)

Dynamic range: 103

  • Flame photometric detector (FPD) 

Temperature setting: Maximum 350℃

Detection limit: ≤2×10-12 g/s (P) ≤4×10-11 g/s (s)

Sample: Methyl parathion

Dynamic range: P 103

Dynamic range: S 102

Heating area

  • In addition to the column box independent heating area, there are six heating areas. 2 injector heating zones, 2 detector heating zones and 2 auxiliary heating zones.